La descarga está en progreso. Por favor, espere

La descarga está en progreso. Por favor, espere

Part No...., Module No....Lesson No

Presentaciones similares


Presentación del tema: "Part No...., Module No....Lesson No"— Transcripción de la presentación:

1 Part No...., Module No....Lesson No
Module title Material de entrenamiento del OIEA sobre Protección Radiológica en radiodiagnóstico y en radiología intervencionista PROTECCIÓN RADIOLÓGICA EN RADIODIAGNÓSTICO Y EN RADIOLOGÍA INTERVENCIONISTA Parte 15.4: Optimización de la protección en radiografía Ejercicio práctico Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

2 Perspectiva general / Objetivos
Part No...., Module No....Lesson No Module title Perspectiva general / Objetivos Materia objeto: control de calidad de un sistema general de radiografía Procedimiento a seguir paso a paso para implantar el ensayo de QC considerado Linealidad y reciprocidad de mA y tiempo de exposición Exactitud del selector de tiempo Interpretación de resultados Explanation or/and additional information Instructions for the lecturer/trainer IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

3 Parte 15.4: Optimización de la protección en radiografía
Part No...., Module No....Lesson No Module title Material de entrenamiento del OIEA sobre Protección Radiológica en radiodiagnóstico y en radiología intervencionista Parte 15.4: Optimización de la protección en radiografía Linealidad y reciprocidad de mA y tiempo de exposición Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

4 ¿Cuál es el equipamiento mínimo necesario?
Equipo electrónico (multímetro) para medir: Tiempo de exposición mA Buenos protocolos

5 Exactitud del selector de tiempo (I)
Método Aproximadamente 70 kVp, 200 mA Medir tiempo a todos los valores comúnmente usados Calcular el error entre tiempo seleccionado y real

6 Exactitud del selector de tiempo (II)
El tiempo se mide normalmente desde el instante en que la salida del tubo de rayos X alcanza el 75 % del máximo hasta que cae por debajo de 75 % Esto puede crear problemas cuando la onda de exposición presenta tiempos largos de subida y caída

7 Exactitud del selector de tiempo (III)
Intensidad de rayos X Tiempo 75 % del máx. Tiempo de exposición medido

8 Exactitud del selector de tiempo (IV)
Análisis El tiempo real debe caer dentro del  5% del tiempo seleccionado, o bien  1 pulso (monofásicos, exposiciones cortas) Frecuencia Aceptación, mantenimiento del selector de tiempo, anual