Mercedes Miñano, Urmila Soldevila Detectores Mercedes Miñano, Urmila Soldevila 27 Marzo 2009
Detectors measured with the ALIBAVA system P-type Detectors: Samples The Mask set has been designed within the CERN RD50 Collaboration (http://cern.ch/rd50) . Microstrip detectors Irradiated with neutrons P-type Detectors: Samples : ATLAS07 miniature microstrip detectors Hamamatsu Sensor Irradiated with protons and neutrons
Fluence (1MeV equivalent neutron /cm2) Wafer Technology + Growing Method Sensors Fluence (1MeV equivalent neutron /cm2) Irradiation CNM sensors: 4 n-on-p FZ 7(-1) 0(2), 1014, 3x1014, 1015, [3x1015, 8x1015] No medidos neutron 17 n-on-n FZ 6 0, 1014, 3x1014, 1015, [3x1015, 8x1015] IV, no signal (volver a medir?) 10 n-on-p MCz 0(2) , 1014, 3x1014, 1015, 3x1015, 8x1015 Se intentó medir un no irradiado Microdescargas El resto no están medidos 19 p-on-n MCz 6 (-1) 0, 1014, 3x1014, 1015, 3x1015, 8x1015 (el no irradiado fue pegado en una placa para tests iniciales del ALIBAVA, datos perdidos) El resto no están medidos Hamamatsu sensors: 4+4 0, 2x1014, 5x1014, 1015 n-on-p FZ proton TOTAL 39 Detectores con microdescargas!! Con este quemamos el chip de una daugther board
Wafer 4: P-type FZ Sensor name Fluence (n/cm2) V md (V) I-V info S08 Non-Irradiated 170 Done S01 1014 250 Vbd ~270V S02 3x1014 270 (a second measure 150V) S03 1015 330 S04 3x1015 - S05 8x1015 Wafer 4: P-type FZ
Wafer 17: N-on-N FZ Sensor name Fluence (n/cm2) V md (V) I-V info S24 Non-Irradiated 230 - S01 1014 390 Vbd ~680V S02 3x1014 400 S03 1015 S04 3x1015 No signal Vbd >750V S05 8x1015 Vbd>750V
Hamamatsu sensors: P-type FZ Detectores en los que también se observan microdescargas.
CNM vs Hamamatsu sensors
Otras cuestiones… El congelador del laboratorio está K.O. Será necesario comprar otro congelador. (Mejor un congelador vertical)